Fast* neutron water kerma and flux measured between
1999. Okt. - 2000. Febr.

No Date Filter#
mm
Circumstances Dose rate
mGy.s-1
Inc. Flux
n.cm-2.s-1
Monitor
cps
1 X.18. 0 Bi
45Pb+10B4C
On B4C filter 4.36 4.05*108 17400
2 X.25. 50 Bi
10B4C
On B4C filter 2.81 2.60*108 9600
3 X.25. 150 Bi
45Pb+10B4C
On B4C filter 1.72 1.62*107 2360
4 X.26. 0 Bi
100 plexi
On B4C filter 2.99 2.77*108 12400
5 X.27. 200 Bp+20Fe
45Pb+10B4C
On B4C filter 0.0233 2.17*106 600
6 I.17. 50 Bi
45Pb+10B4C
On B4C filter 1.32 1.20*108 9000

# In the first line the internal, in the second line the external filters are given in mm.
* Neutron energy > than 100 keV.

Thermal neutron water kerma and flux measured between
1999. Okt. – 2000. Febr.

No Date Filter
mm
Circumstances Dose rate
mGy.s-1
Inc. Flux
n.cm-2.s-1
Refl. Flux
n.cm-2.s-1
Monitor
cps
1 X.18. 0 Bi
45Pb+10B4C
10 cm far from
B4C filter
0.004 1.75*105 - 17400
2 X.19. 0 Bi
45Pb+10B4
C
Between B4C
filter and 3 cm
water
0.006 2.98*105 2.95*105 16900
3 X.19. 0 Bi
45Pb
On B4C filter 1.176 1.76*106 - -
4 X.25. 150 Bi
45Pb+10B4C
Between B4C
filter and 2 cm
water
0.0004 9.57*103 5.18*103 2360
5 X.26. 0 Bi
100 plexi
On B4C filter 0.016 4.39*105 6.34*104 12400
6 X.26. 0 Bi
45Pb+100 plexi
On B4C filter 0.014 2.61*105 8.64*103 9600
7 XI.15. 50 Bi
45Pb+10B4C
On B4C filter 0.0017 6.20*105 2.93*104 9000

Notes
Dose: incident + reflected together.


Chapter 4/ page 13 of 17